Practical radiation oncology 2021 Feb 23
Transition From Manual to Automated Planning and Delivery of Volumetric Modulated Arc Therapy Stereotactic Radiosurgery: Clinical, Dosimetric, and Quality Assurance Results.   
ABSTRACT
PURPOSE
Properly planned single isocenter volumetric modulated arc therapy (VMAT) radiosurgery plans exhibit high quality and efficiency. We report here the largest clinical experience to date, to our knowledge, comparing manual planning with a new automated platform designed to standardize and simplify radiosurgery planning and delivery processes.
METHODS
We treated 693 patients with single isocenter VMAT radiosurgical plans generated by either our conventional manual (mVMAT) or a recently implemented automated (HyperArc) technique. All plans targeted the gross tumor volume without margin. Radiochromic film was used for patient-specific quality assurance (PSQA). We evaluated local control and toxicity data for a subgroup of 107 patients having 377 metastatic tumors that were treated with HyperArc.
RESULTS
The median Radiation Therapy Oncology Group (RTOG) conformity index was 1.14 and was not different between the 2 techniques. The median Paddick gradient index was 5.42 for HyperArc versus 7.09 for mVMAT (P < .001). The median mean brain doses were 4.6% and 5.1% for HyperArc and mVMAT, respectively (P = .04). The PSQA for both techniques met clinical criteria, but 97% of the HyperArc plans satisfied the gamma tolerance limit recommended by the American Association of Physicists in Medicine Task Group No. 218, compared with 94% of the mVMAT plans (P = .02). The median treatment-planning times were not significantly different. The median treatment times were 10.5 and 11.4 minutes for HyperArc and mVMAT, respectively (P < .001). The Kaplan-Meier estimate of local control was 90.1% at 1 year.
CONCLUSIONS
HyperArc produces high-quality radiosurgical plans that are at least as good as mVMAT plans created by an expert manual planner with easier planning and more efficient delivery workflow. A less experienced planner can produce very high-quality radiosurgical plans even for patients with more than 10 targets. The use of a single-isocenter technique for multiple targets with no PTV margin did not compromise clinical outcomes, and 1-year local control for treated targets remained congruent with historical series.