What is the maximum V20 on ipsilateral lung that can be safely accepted for 3 or 4-field breast plans?

V20 of 30% can be hard to attain if IMs are being treated.
3 Answers
Mednet Member
Mednet MemberInvited Expert
Radiation Oncology · University of Texas MD Anderson Cancer Center
Answered on

We routinely treat the IM nodes so my comments reflect this practice:

1. I shoot for a mean ipsilateral lung V20 below 35%. This is achievable in most, though not all, plans.

2. The biggest driver of ipsilateral lung V20 is the amount of lung in the SCV field.

3. While it is tempting to raise the ma...

Join for free or sign in to see the full answer

Mednet Member
Mednet Member
Radiation Oncology · Rush University Medical Center
Answered on

I believe that the we sometimes forget about the other risk of lung radiation - secondary lung cancers. An older NSABP study and the following (Risk of second primary lung cancer in women after radiotherapy for breast cancer. Grantzau T; Thomsen MS; Vaeth M; Overgaard J. Radiotherapy & Oncology. 111...

Join for free or sign in to see the full answer

Mednet Member
Mednet MemberInvited Expert
Radiation Oncology · Varian Medical Systems/Allegheny health network
Answered on

We try to keep the V20 below 30 percent for the ipsilateral lung and use different techniques to achieve that goal including matching electrons, an electron and photon combination, or photons alone.

Join for free or sign in to see the full answer