How do you determine the appropriate toric IOL when there is a discrepancy in axis or cylinder power between optical biometry and corneal tomography during preoperative evaluation?

3
4 Answers
Mednet Member
Mednet Member
Ophthalmology · NYEEI Mt Sinai

Always difficult. Always re-measure when discrepancies occur. For axis discrepancies, Auto Ks, topography, biometry, and past refraction. Past refraction is where the patient lived their entire life before their cataract surgery. Which of these tests aligns the most? Past refraction, especially pre-...

Register or Sign In to see full answer

Mednet Member
Mednet Member
Ophthalmology · Whitten Laser Eye

In my experience, the most common causes of discrepancies between preop measurements are ocular surface disease or ABMD, as others have mentioned. I like to see a 10° or less difference across my devices. More than that, then I take a hard second look, treat the ocular surface, repeat measurements, ...

Register or Sign In to see full answer

Mednet Member
Mednet Member
Ophthalmology · UNC

I recommend double-checking both measurements; poor alignment or dry eye can cause a discrepancy.

If the discrepancy is still high (over half a diopter), the patient probably presents with an irregular astigmatism, and a toric IOL is not recommended.

If they are slightly different, I usually use the...

Register or Sign In to see full answer

Mednet Member
Mednet Member
Ophthalmology · Center For Advanced Eye Care

Manual Ks are the best.

If the topo rings are irregular, look for ABMD.

It is often very helpful to focus on DES and have them return for repeats.

Register or Sign In to see full answer